Keywords: Surface dose; in vivo dosimetry
There are essentially three processes that occur on
the skin surface: electron backscatter within the phantom,
the formation of secondary electrons in the irradiated
material, and the formation of contaminant
electrons in the head of the therapy device.[
Megavoltage photon beams have a skin-sparing
effect because the surface dose is usually much lower
than the maximum dose that occurs under the skin.[
2. In Vivo Dosimetry And Measurement Points
There are many reasons for using the in vivo dosimetry
(IVD) method in radiotherapy. The most ambitious
application of IVD is to control the dose in the
target volume in order to verify that the applied beam
is correct. By placing the detectors in natural body
cavities, such as the esophagus, rectum or vagina, the
target organ dose can be obtained. However, placement
outside these regions will unlikely be able to measure
the target organ dose accurately. In recent years, by
means of wireless IVD systems, detectors can be placed
in designated areas of the body from which the target
volume can be determined. Due to the high cost of
these systems, however, their use is limited.
The determination of the skin dose represents another
possible application of IVD. To measure this, the
dose derived from the signal of detectors on the skin
surface is compared with theoretical values, and the
dose is subsequently calculated by TPSs. However, the
accuracy of the skin surface dose calculation is controversial,
and TPSs cannot correctly calculate the surface
dose.[
3. Measurement Points Used For In Vivo Dosimetry
3.1. High-Energy Photon Beams (Dentrance, Dexit,
Dsurface, and Dtarget)
For a single high-energy photon beam, the dose at
the depth representing the maximum dose (dmax) from
the entrance surface is called the "entrance dose" (Dentrance),
whereas the dmax from the exit surface is called
the "exit dose" (Dexit). The entrance and exit dose points
are symmetrical with respect to the midline. From the
entrance point of the medium irradiated by a single
photon beam, the dose slowly rises from a low value on
the surface to the maximum Dentrance dose, which depends
on factors such as energy, collimator opening, and
source skin distance.[
In general, the increase in dose, which is a function
of depth from the surface to dmax, starts immediately
below the surface with the slope decreasing at deeper
distances and finally reaching a plateau as it approaches
the dmax. This means that for Dentrance measurements,
adequate material must be present around and in front
of the detector placed on the skin to make the measurements
repeatable. To limit the effect of head-scattered
electrons on the Dentrance, a build-up cap must be used
without any accessories, at a size that achieves complete
build-up in the smallest collimator aperture.
The most accurate measurements are obtained when
the detector is not within high-dose gradient regions.
This can be achieved with the correct build-up thickness.
The build-down region, which is related to the lack
of backscatter radiation from the air behind the patient,
is located at the exit side of the patient. Photons
and secondary electrons are responsible for this lack of
backscatter, which causes the dose to decrease only a few
millimeters beyond the exit surface. To achieve complete
electron backscatter for Dexit measurements at dmax
from the exit surface, it is important to use enough cover material over and around the detector. The determination
of the target dose is more complicated than the dose
measurements of the other points used in IVD. To identify
and correct possible errors, each beam contributing
to the target dose should be individually controlled, at
least during the first fraction of treatment. In addition,
it is important to check whether the treatment is repeatable
over the following fractions.
In most cases, the in vivo measurement of both the
exit and entrance doses is performed simultaneously.
Extreme care must be taken when placing the exit detector
because of the possible shadowing effect of the
entrance detector.
These techniques are valid for conventional radiotherapy.
Today, by means of new-generation radiotherapy
technologies, any point of the target can be obtained
using 3D TPSs. Dose control can be achieved by IVD if
the target is in a body cavity. Otherwise, the other points
used in the IVD can be measured and compared with
the TPS to check the accuracy of patient treatment.
3.2. High-Energy Electrons (Dentrance, Dexit, Dsurface,
and Dtarget)
Because the target volume is most often electron irradiation
of the patient"s skin, the Dsurface and Dtarget
can be related to each other. A single electron field is
generally used in clinical applications of electron irradiation.
However, additional electron fields can be
used, particularly for breast irradiation or total-skin irradiation.
Moreover, electron beams are easily affected
by a lack of homogeneity. If the target is located around
or inside a non-homogenous region, over- or underdosage
problems may occur. Therefore, IVD for electron
beams requires great care. According to the ICRU
(1985), Dsurface is defined as the dose at 0.07 mm
beneath the surface. Thin detectors, such as TL chips,
should be used for measurement because they do not
require any build-up material. They can be covered by
a piece of thin paper and stuck onto the patient"s skin.
According to the ICRU recommendations, the target
dose (Dtarget) is defined as dmax, which is the depth
at which Dentrance also occurs (Fig.
4. Detectors Used in In Vivo Dosimetry
Diodes, metal-oxide-semiconductor field-effect
transistors (MOSFETs), plastic scintillation detectors (PSDs), and electronic portal imaging devices (EPIDs)
are real-time detectors used in IVD. These detectors
are called real-time detectors because they provide a
real-time dose response during the treatment.
Thermoluminescent dosimeters (TLDs), optically
stimulated luminescent dosimeters (OSLDs), implantable
MOSFET detectors, radiophotoluminescent dosimeters
(RPLDs), and film (radiographic and radiochromic)
are passive detectors. These detectors cannot
provide immediate dose readings during treatment
because the measurements need to be converted into a
dose. While TLD, OSL, RPLD, and implanted semiconductor
detectors allow point dose measurement, films
provide 2D dose information.[
4.1 Real-Time Detectors Used in In Vivo Dosimetry
Silicon diodes
The dose response of silicon diodes can show
slight variation. The cable used to provide the connection
between the diodes placed on the patient and the
electrometer can be considered a disadvantage of this
type of detector. However, new-generation systems use
wireless technology. The diode sensitivity may change
after accumulating high doses at adequate levels; therefore,
they need to be calibrated regularly throughout
the period of clinical use.[
Metal-oxide-semiconductor field-effect transistors
The detector has two different surfaces: round and
flat. Generally, when the skin dose of the patient is being
investigated, the MOSFET detector is adjusted so
that the flat side faces the skin and the round side (epoxy
side) faces the beam source. The water-equivalent
measurement depth of the MOSFET detector is approximately
0.8 mm for the rounded portion and 1.8
mm for the flat portion. Over the past 5 years, MOSFETs
have been used in in vivo and phantom irradiations,
including for routine patient-dose verification,
brachytherapy applications, total body irradiation
(TBI), intensity-modulated radiotherapy (IMRT), intraoperative
radiotherapy, and radiosurgery. Depending
on the application, they can be used with or without
build-up caps.[
The advantages of MOSFET dosimetry include
their small dimensions, reproducibility, consistent results,
measurements independent of dose rate, and
negligible angle dependence.
One of the disadvantages of MOSFET dosimetry
is their temperature and energy dependence. In addition,
the dose response is reduced as a cumulative dose
function, and the dose slowly decreases after irradiation.[
Plastic scintillation detectors
Once calibrated, PSDs do not need correction factors,
which are used by other detectors to convert the
dosimeter readings into an absorbed dose. They have
excellent spatial resolution due to their small size. For
IVD using PSD, the difference between the measured
and expected doses has been reported to be<1%.12
Unfortunately, because they are not currently available
on the market, they do not yet have a wide range of
clinical applications.[
Electronic portal imaging devices
The disadvantages of using an EPID as a dosimeter
include the fact that the flat-panel imaging material is
highly sensitive to low-energy photon applications that
are affected by water non-equivalent material. For example,
continuing signal reception after radiation interruption
is an example of a problem related to the use
of an a-Si EPID (ghosting). Consequently, the EPID
response is not linear to the dose.
There are two different approaches for using EPID
in IVD. In the first approach, a portal dose image of the
patient at the position of the EPID is predicted using
the planning CT data; these data can then be compared.
With this technique, it is not clear how dose differences
in the EPID plane are related to dose differences in the
patient. For this reason, a method known as "backprojection"
has been explored by several groups. In this
method, the patient dose distribution is derived from
a measured portal dose image. The method applied by
Piermattei et al.[
4.2 Passive Detectors Used for In Vivo Dosimetry
Thermoluminescent dosimeters
After irradiation, thermoluminescence signals decrease
with time due to the spontaneous emission of
light at room temperature (fading). The characteristics
that make TLDs suitable for IVD include their ability
to be used repeatedly after resetting, their ability to be
placed in the patient"s body cavities after being properly
wrapped, and their ability to be used with a small
correction factor. The use of TLD dosimetry requires
great attention because the effective measurement distance
changes according to the TLD material.
Optically stimulated luminescent dosimeters
Optically stimulated luminescent nanodots can
be used in three different configurations: closed,
open-down, and open-up. The effective measurement
depths change depending on the configuration of the
nanodots used. The effective measurement depths for
closed, open-down, and open-up have been reported
to be 0.85, 0.55, and 0.35 mm, respectively.[
The use of light for signal reading provides superior
control over reading of the dosimeter with a quick on-off capability. The applications of OSLDs include output
verification as well as phantom or in vivo surface
dosimetry.
Although the basic properties of OSLD obtained
using aluminum oxide have been investigated, there
are still unanswered questions regarding the clinical
performance of commercial OSLDs in radiotherapy.
[15] OSLDs with different configurations and reading
systems are shown in Figure
Radiophotoluminescent dosimeters
Film (radiographic and radiochromic)
Radiochromic films are a new type of film used
in radiotherapy dosimetry. Gafchromic films are the
most widely used. Radiochromic films contain a special
polymerized dye and emit a different color in the
radiation-exposed region. The polymer absorbs the
light, which is transmitted through the film and can be
measured using a suitable densitometer, such as that in
radiographic films. Flatbed color scanners with 48-bit
image depths are usually used for reading. New-generation
films exhibit polarization effects depending on the
orientation of the film. For this reason, it is necessary to
pay close attention to the placement of the films when
taking measurements.
The responses of both radiographic and radiochromic
films in the megavoltage range are stable and unchanged.
The greatest advantage of radiochromic films
over radiographic films is that they do not have any energy
dependence, even at keV levels over a wide energy
range.
Unlike TLD, OSL, and RPLD, films can give the user
2D dosing information and can permanently record
radiation. The use of radiochromic films in dosimetry
applications is rapidly increasing and is often preferred
in IVD applications for skin dose measurements, total
skin electron irradiation, and TBI measurements.[
Gafchromic films are more practical dosimetric systems
because they are not sensitive to light, water equivalent,
independent from dose rate and dose fraction,
do not require a dark room and bathing, unaffected by
watery or bloody environments; in addition, films can
be cut into various sizes.[
Implantable semiconductor detectors
In vivo is a Latin word meaning "within the living
thing," whereby living organisms are used for various
purposes. In terms of radiotherapy, it means the measurement
of the dose that reaches the patient during treatment, as opposed to the in vitro or ex vivo measurements
of phantoms before or after treatment.
First, the measurement points used in IVD should be
discussed. Different layers of skin, including basal and
dermal layers, have varying depths. These depths differ
among the layers, from patient to patient, and even
where they are located in different patients. This makes
skin dosimetry even more complicated. Despite the uncertainty
with regard to depth values, the International
Commission on Radiological Protection and International
Commission on Radiation Units and Measurements
(ICRU, 1985) have made some recommendations.
Accordingly, the recommended depth for measuring the
skin dose is 0.07 mm, which is the depth of the basal
layer.[
High-energy electron beams have a uniform dose
distribution on the surface; however, after a specific
depth, a rapid dose decrease is observed. The penetration capability of electron beams is limited. Due to
these characteristics, determining an exit dose for electron
irradiations is not possible, particularly for thicker
anatomical structures.
The detectors used in IVD can be classified into two
categories: real-time and passive. Both types of detectors
require calibration. Calibration is usually performed
by comparing the dose response with a calibrated
ion chamber in a specific radiation field. Most
detectors used in IVD have energy and dose responses.
For this reason, considering that the actual irradiation
conditions differ from the calibration conditions, some
correction factors should be used when determining
their dose response.
Silicon diodes were first used in the early 1980s. Figure
MOSFETs are detectors that include a p-type silicon
semiconductor layer. MOSFETs are miniature silicon
transistors with high spatial resolution that do not disrupt the beam too much due to its small size. MOSFETs
measure the threshold voltage, which is a linear function
of the absorbed dose. They require a bias voltage
during irradiation, and their lifetime is limited.[
PSDs are promising for IVD as well as quality assurance
(QA) applications due to their favorable dosimetric characteristics, such as water equivalency, energy
independence, dose linearity, and radiation damage
resistance.[
EPIDs usually provide megavoltage portal images
of the patient in a digital format that can be used to
identify set-up errors during treatment. They can also
be used for dosimetric purposes. They are useful for
dose-control purposes, particularly for advanced treatment
techniques, such as IMRT and VMAT. EPIDs
are a promising system for IVD because they are not
interventional and can provide 2D and 3D dosage information.
Indeed, EPIDs with a flat-panel detector
system based on amorphous silicon (a-Si)-photodiode
technology are frequently used. The response of an a-Si
EPID is independent of the dose rate and is approximately linear to the integrated dose.
The use of TLDs, a type of passive detector used in IVD
measurements, is based on years ago in radiotherapy.
TLD-100 (LiF:Mg,Ti) is commonly used for TLD in
IVD. TLDs can be found in many different forms, including
powders, chips, rods, or ribbons. TLDs can be
used more than once, but they must be exposed to high
heat before use. Furthermore, TLDs should be separated
into groups before they are used in dosimetric applications
by combining TLDs with similar dose values
after irradiation. One or two TLDs from these groups
can be allocated for calibration purposes. These TLDs
are irradiated under the measurement conditions from
which the dose efficiency of the device is obtained. The numerical value, corresponding to the radiation response
of the TLD, is translated into the dose. For the
calibration of TLDs, the use of the same feature beam
that was used during patient treatment is recommended.[
OSLDs comprise carbon-doped aluminum oxide
(Al2O3:C). While their application in radiotherapy is
recent, they have been used for many years in radiation
protection. OSLDs are similar to TLDs in terms
of their dosimetric properties. The electrons released
by exposure to ionizing radiation are caught in energy
traps contained within the forbidden energy bands of
crystal defects. This creates electron-hole pairs in the
lattice. Exposing the substance to visible light stimulates
the trapped electrons and causes them to recombine
by emitting an optical photon. The optical photon
fluence is proportional to the dose.[
RPLDs were historically used for personal dosimetry in
the 1950s and 1960s but are now used in radiotherapy
measurements.[
Film (radiographic and radiochromic) is a passive detector
used in IVD. There are two different types of films: radiographic and radiochromic. Radiographic
films comprise a transparent substrate and an emulsion
layer coated on either side of the substrate. There are
a large number of AgBr crystals in the gelatin within
the emulsion layer. The AgBr in the emulsion layer is
sensitive to light and X-rays. Some changes occur in the
AgBr crystals exposed to X-ray photons. A latent image,
which is not visible to the naked eye, appears on
the film. This image is made visible only by a series of
chemical processes. Beam transmission is a function of
film opacity and is measured as optical density (OD)
using a densitometer. The OD is a function of the dose.
Implantable semiconductor detectors were designed to
measure a patient"s daily dose when undergoing radiotherapy. Implantable detectors comprise a dual MOSFET
detector, a data acquisition chip, a microprocessor,
and a copper bobbin. All parts of the detector are
encapsulated in a glass tube. The system has a portable
telemetric reader attached to the dosimetry antenna,
which provides power to the dosimeter, enabling data
transfer. The dosimeter remains passive during irradiation
and opens to measure the threshold voltage only
after treatment has ended. Their dose repeatability is
5% or better. Because they are not used very often in
clinical practice due to reasons including surgical implantation
and permanent implantation, they are not
likely to be further developed.
Disclosures Statement
The author declare no conflicts of interest.
Peer-review: Externally peer-reviewed.
Conflict of Interest: None declared.