METHODS
The CK plans of ten patients with intracranial tumors were selected for this study. The plans were transferred
to computed tomography images of the anthropomorphic head phantom. The phantom containing
the film was irradiated using 6D skull tracking method. The dose distributions calculated by TPS
were compared with those measured by film using gamma analysis method.
RESULTS
In the gamma-index method, the average passing rate for the 3%/3 mm criterion was found to be
93.9±4.8%. The gamma passing rate for Iris-collimator-based plans yielded slightly lower than that for
fixed-collimator-based plans.
CONCLUSION
In conclusion, this study indicates that Gafchromic EBT2 film is a utilizable dosimeter for quality assurance
of the CK treatment plans. It is recommended to use the EBT2 film for verification of the CK
treatment plans before delivery to the patient.
Keywords: CyberKnife; film dosimetry; gamma evaluation; quality assurance
The CyberKnife (CK) is a unique system comprising
of a robotic-based linear accelerator. In the CK
treatments, for the rapid fall-off of dose between the
target and normal tissues, 150-300 non-isocentric
beams from different node positions are delivered to rangthe
target with a compact linear accelerator mounted
on a robotic arm. Throughout the treatment delivery,
tumor position is monitored through an imaging system.
The CK treatment delivery is very complicated
due to all these factors, and hence, strict quality assurance
(QA) tests are required.[
The periodic machine QA guidance has been explained
in the report from AAPM Task Group 135.
In this report, it is recommended that the automated-
quality-assurance (AQA) test and the end-to-end
(E2E) test be routinely carried out for mechanical accuracy.
The implementation processes of both tests
are described in detail. The AQA test is used for determining
isocentric targeting accuracy. The E2E test
is used to check the overall targeting accuracy of the
CK system. They cannot be used for determination of
dose accuracy for treatment plans of non-isocentric
patient. It is also recommended that the delivery quality
of assurance (DQA) test, which assesses both spatial
and dosimetric accuracy of delivery, is performed
as a part of the machine commissioning and monthly
QA using film or detectors.[
The radiochromic film is an ideal dosimeter to obtain
dose distributions in high dose gradient regions
due to its high spatial resolution and low spectral sensitivity.
Films provide two-dimensional (2D) dose maps
of treatment plans. Regarding spatial resolution, they
are superior to other 2D radiation detectors.[
The purpose of this study was to verify the CK dose
distributions generated for patients with intracranial
tumors using Gafchromic EBT2 film.
2.1. Film Calibration
First, a calibration curve was created before using
the films to get a dose map related to the plans. Gafchromic
films were cut into 2×2 cm2 pieces, and they
were placed between the solid water slab phantoms at a
depth of 5 cm. The source to film distance was 100 cm.
Films were oriented perpendicular to the central axis
of the beam, and they were irradiated with doses rangthe ing from 1 to 800 cGy at a 10×10 cm2 field size using a
photon energy of 6 MV. An unexposed film was used
to assess the background. All films were scanned using
a flatbed scanner (Epson 10000XL America Inc., Long
Beach, CA) on the following day to provide maximum
post-irradiation darkening.[
2.2. 2D Dose Distribution Measurements
In this study, Gafchromic EBT2 (Lot no.: A04141003BB)
(ISP, International Specialty Products, ABD) film was
used. The EBT2 films comprise a single active layer
with a nominal thickness of 30 µm. The active part of
the film is asymmetrically located between the layers.
The substrate is made of polyester with thickness of 175
µm. There are three protective sheets with a total thickness
of 80 µm over the active layer. This non-symmetrical
layer configuration causes different film responses
for the same scanning orientation when different sides
of the films are facing the scanner. The measurable dose
range of the EBT2 film was 1-800 cGy.[
The anthropomorphic head phantom was used for film
dosimetry. The phantom mimics the bone structures
and soft tissues of an adult human head. A phantom
with six RW3 mini plates (sized 6×9 cm2) each of which
consists of 1 cm thickness was prepared. Gafchromic
EBT2 film was put in the center of this phantom. Then,
the prepared mini phantom was placed in the special
cavity of the anthropomorphic head phantom (Fig.
For other five plans based on Iris collimator, the
gamma pass rates using same criteria are 72.9±10.0%,
91.9±6.6%, 97.3±3.5%, and 99.0±1.8%, respectively.
The percentage gamma-index value for Iris-collimatorbased
plans yielded slightly lower than that for fixedcollimator-
based plans (Fig.
Gafchromic films have been frequently used for
patient-specific QA. Kairn et al. reported that Gafchromic
EBT2 could be used for QA of complex
IMRT treatment because of its high spatial resolution
and minimal angular dependence of response.
[
There is no study on the comparison of QA of the
treatment plans created using different collimator
types. In addition, there is no guideline about QA of
treatment delivery for Iris-collimator-based plans. In
our study, we compared the QA results of plans generated
using Iris and fixed collimator. Gamma-index
pass rates using the 2%/2 mm, 3%/3 mm, 4%/4 mm,
and 5%/5 mm criteria for plans based on Iris collimator
were found slightly lower than those for plans
based on fixed collimator. Physical characteristics of
Iris collimator that has two collimator banks with six
leaves each may affect the results.
Peer-review: Externally peer-reviewed.
Conflict of Interest: None declared.
Financial Support: None declared.
Authorship contributions: Concept - C.K., G.K.; Design - C.K., G.K.; Supervision - C.K., H.B., G.K.; Materials - C.K.; Data collection &/or processing - C.K., U.A., N.D.K.; Analysis and/or interpretation - C.K., H.B., G.K.; Literature search - C.K., G.K.; Writing - C.K.; Critical review - H.B., G.K.