METHODS
In this study, a tumor volume was determined in the middle esophagus in the digital phantom. By targeting
this tumor volume, a three-dimensional conformal radiotherapy treatment plan was created and
dose-volume histograms (DVH) were compared according to the Pencil Beam, Collapsed Cone,and
Monte Carlo algorithms. The total dose was determined as 5040cGy (1.8 Gy/fraction). In DVH; mean
planning target volume dose was evaluated as D50, D98, D2; mean dose for the heart as V5, V30; mean
dose for lung as V5, V20; and also the maximum dose (Dmax) for the spinal cord and homogeneity
index were assessed. A total of 18 plans created at the same energy levels (6 and 18MV) and angles (3, 4,
and 5 fields) were compared using these three different algorithms.
RESULTS
Different algorithms created significant differences with the same energy and same field angles as we expected.
Especially when considered in terms of normal tissues, the remarkable difference was in the heart
(Dmean, V5), lung (Dmean, V5, V20), and spinal cord Dmax values. There were also differences in algorithms
between PTV dose values. We found that with the increase in the energy level and field, the dose
differences between algorithms significantly reduced.
CONCLUSION
Variations between algorithms that may occur due to the difference in density between tissues in the thoracic
region should be taken into consideration.
Keywords: Collapsed cone; digital phantom; esophageal cancer; monte carlo; pencil beam
With the occurrence of different algorithms in TPSs
in the past decade in some studies, the radiobiological
and dosimetric impact of Pencil Beam (PB) versus
Monte Carlo (MC) dose algorithms were evaluated.
[
In our study, the esophagus was preferred due to its
location in a heterogeneous region and PB, Collapsed
Cone (CC), and MC algorithms are compared dosimetrically.
The aim is to show the differences between
algorithms in a heterogeneous environment, and consequentially
to emphasize that it may have a significant
impact on future treatment plans.
The reason for choosing the middle esophagus as the primary tumor site is that this area is highly heterogeneous due to the presence of lung, heart, costa, vertebra, and spinal cord. In addition to PTV, critical organs such as lungs, spinal cord, and heart were contoured. The plans were created using the same target volume and critical structure contours in all plans for an effective comparison. In particular, we would like to point out that proving the accuracy of these three algorithms is not among our aims. The planning was prepared by protecting the dose constant as if the daily fraction dose was delivered to the real patient with 28 fractions of 180cGy. Segments and wedge were not used in the plans. It was not tried to choose ideal angles to protect the organs. We, the physicians, determined the angles (120, 90, and 72), fields (3, 4, and 5), energies (6-18MV), and then a total of 18 plans were investigated, according to PB, CC, and MC algorithms.
Ninety-five percent of PTV was covered by prescription isodose. On the (DVH), we evaluated D50, D98, D2, and mean doses for the PTV; mean dose, V5, V30 for the heart; mean dose, V5, V20 for lung; and Dmax for the spinal cord. In addition, we also compared the homogeneity index (HI) which evaluates the homogeneity of the dose distribution of the PTV. HI was calculated according to ICRU Report62 ([D2- D98]/D50). Dose differences between PB, CC, and MC algorithms were examined.
Monaco TPS
Monaco TPS developed by Elekta Company was used
in this study. There are manual and automatic contouring
tools in the contouring part where the target volumes
and OARs are drawn and their boundaries are
determined. In the planning section, there are tools by
which parameters such as the beam"s gantry angle and
Multi-Leaf Collimator can be manipulated. By changing
and editing these parameters, OARs are preserved
while giving the desired dose to the target volume.
Monaco has the capability to calculate modalities such
as photons, electrons, and protons beams.
For these calculations, the PB, CC, and MC algorithms can be used by the user.
When the user starts to plan, the user chooses the algorithm to be calculated after selecting the number of beams, gantry angles of the beams, beam energy, etc.In three-dimensional conformal radiotherapy (3D-CRT) plans, the user defines a prescribed dose for the target region then the calculation starts. After the calculation is completed some hot and cold dose regions could appear in the treated volume. The user needs to adjust them manually. In IMRT and volumetric modulated arc therapy (VMAT) plans, this is done by Monaco according to the optimization parameters created by the user. The user adds the drawn structures to the optimization at this stage and defines the cost functions is available in the system to the optimization according to the type of these structures (Target, OAR). By manipulating the dose distribution with these parameters, while taking the maximum dose of the target volumes, it is ensured that the OARs are affected to the minimum from this dose. Monaco TPS has two optimization modes named "Pareto" and "Constrained." In the Pareto mode, the cost functions that aim to deliver the prescribed dose to the target volume are more powerful than the cost functions that try to keep the dose received by OARs within a certain limit. In other words, priority is given to target volumes in this mode. In the other mode, constrained, the situation is the opposite, that is, priority belongs to OARs. The constrained mode is used in all plans in this study.
Mathematical Digital Phantom
Various mathematical and statistical methods such as
MC are used to calculate the behavior of radiation in the
human body. Various software uses these methods and
it has been ensured that radiation transport calculations
such as dose and flux distribution are performed in computer
environment. This software simulates the interaction
of radiation with matter using cross section data.
Cross-section data include the probability distribution
of various events that are likely to occur as a result of the
interaction by defining each isotope depending on the
type and energy of the radiation. For this reason, to simulate
the interaction of radiation with matter, the environment
must be described in detail at the atomic scale.
Various mathematical models and phantoms have
been produced to perform radiation transport calculations
in computer environment. In the past, these
models have been defined using simple geometric
shapes such as square, rectangular, ellipse, and cylinder
medical internal radiation dose (Fig.
The ICRP Reference Male Phantom is produced as
a result of the creation of the human body using voxels
of 2.137 mm × 2.137 mm × 8.0 mm volume. Each of
these small volume voxels are defined using different
material contents at atomic scale. Thus, a 176 cm tall,
73 kg male human model was defined using 1.9 million
voxels. Detailed information about reference phantoms
is given in the report numbered 110 of the ICRP.[
Dose Calculation Algorithms
PB Algorithm
In other words, PB algorithm is not capable at calculation
of the lateral variation of the beam that causes uncertainties.
CC Algorithm
The algorithm uses an approximation where all
energy within a given solid angle will be transported
along a line. The choice of the dose calculation algorithm
can have a huge impact on a treatment plan for a
particular treatment case.[
MC Algorithm
The Elekta (MONACO/MC) is based on a model
using the virtual energy flux model. Dose distribution
within the patient is determined by X ray voxel MC calculation.[
Statistical Analysis
In this study, the algorithms PB, CC, and MC are used
in the calculation of the radiation distribution.
The PB algorithm uses a pre-calculated sample photon
beam to calculate the dose distribution in water. The algorithm
recalculates the dose at different locations using
the intensity distribution along the photon moving path.
The PB creates a dose distribution by integrating it at patient"s
surface to account for changes in primary intensity
and changing the shape of the beam by the effect of depth
and tissue density of the material. The PB algorithm does
not take into account changes in lateral scattering effects.
The algorithm is capable of the calculating the effects of
patient heterogeneities on both primary and secondary
scattered radiation. Naturally, it can take into account
dose distributions in areas with high electron density
variation, such as tissue-air, and tissue-bone.
With the MC method, each photon history (sample)
is calculated one by one for a sufficiently high number
of primary photons and generated secondaries.
Many variance reduction techniques are used to speed
up the calculation and decrease the complexity. In this
method, a particle history is reused and scaled according
to density along its new trajectory. The expected
errors correspond to the expansion of the use of applied
variance reduction methods. This is used as input
data for dose calculations and is itself calculated from a
number of air dose profiles.[
A two-way analysis of variance (ANOVA) was used
to test the differences between different algorithms.
All analyses were conducted in SPSS v. 23. Differences
were reported to be significant at p≤0.05.
In terms of PTV, the lowest minimum dose
(Dmin) and Dmax values were calculated in the CC
algorithm. We know that Dmin values are important
to assess target volume coverage. Considering that
higher Dmin improve the target volume coverage,
MC was the best (high) algorithm because Dmin in
PTV was better with the MC algorithm. However, we
found very high hot dose values when we evaluated
MC algorithm in terms of Dmax. Dmax values occurring
at different angles and energies are shown in
Figures
Considering the MC and PB algorithms of the D2 parameter, which expresses the hot spot formed in our study, we observed that there were significant differences in 6 MV energies. While D2 values in 6 MV energy were 106.9%, 106.1%, and 105.7% for PB algorithm for 3, 4, and 5 field planning respectively, our data for MC algorithm were quite high (111.1%, 110.1%, and 108.9%, respectively).We found that the dose difference between the algorithms decreased with the increase in the energy value. In terms of field number, four field doses were lower than 3 and 5 fields (p<0.05).
In our study, the results in terms of OARs; the max dose value that is important for the spinal cord in the serial organ category was the highest in the MC algorithm and the lowest in the PB algorithm. Considering Dmean values for the heart and lung, which are in the parallel organ category, they were the lowest in the PB algorithm and highest in the MC algorithm. In addition, heart V5 and V30 values were the lowest in the PB algorithm, and highest in the MC algorithm. V5, V10, and V20 values for the whole lung were lowest in the PB algorithm and highest in the MC algorithm.
In terms of HI, no association was found between the algorithms. However, HI values for PB algorithm were the lowest (best homogeneity).
In our study, the most significant difference between
the algorithms occurred in the heart (Dmean, V5), lung
(Dmean, V5, and V20), PTV (D98), and spinal cord
(Dmax) (Table
Thoracic region demonstrates differences between
algorithms in planning due to the tissues with different
densities it contains. Some planning systems require
limitations in clinical use because they ignore electron
and photon scattering in heterogeneous environments.
Especially in tumors such as head and neck and lung cancer, dose calculations do not give accurate results.
With the technological advances, efforts to improve the
dose calculation algorithms used in clinical RT TPSs
have increased. Although PB-based algorithms give acceptable
results in dose calculations in regions with homogeneous
tissue density, some limitations occurred in
the utilization of this algorithm, as it does not correctly
shape the distribution of secondary electrons formed in
heterogeneous tissue density areas.[
Kry et al.[
In another study, they noted that PB and MC algorithms
are compatible in terms of target volume, but MC
algorithm is superior to PB algorithm in a heterogeneous
medium or for critical organ dose calculations.[
Adıgül et al.[
As we know, the dose distribution in PTV must be
homogeneous and an excess of up to + 7% isodoses is
permitted in routine clinical applications in terms of
hot spot (D2). When we examined our study in this
respect, we determined that the hot spots formed in the
MC algorithm are at an unacceptable level, especially
in plans with 6MV energy.
Koeck et al.[
In summary, in our study, we tried to analyze the
effects of different algorithms in the heterogeneous
lung tissue, at the angles and energies, we determined.
The differences between PB and MC algorithms were
significantly greater than the differences between CC
and PB algorithms. Based on the data, according to the
Quantitative Analyses of Normal Tissue Effects in the
Clinic (Ouantec), the algorithm used in dose calculation
should also be considered in terms of both PTV
and OARs. The subject is important in terms of heterogeneity
calculations and clinical applications.
Acknowledgment: We would like to thank the medical physicist Çağrı Yazgan, Orkun Kireççi, Ali Yeşil and Taylan Tuğrul for their support in planning.
Peer-review: Externally peer-reviewed.
Conflict of Interest: The authors have no conflicts of interest to declare.
Ethics Committee Approval: This research is a digital phantom study. Patient data were not used in this study (It was done virtually over a computer system).
Financial Support: The authors declared that this study has received no financial support.
Authorship contributions: Concept - T.B.; Design - T.B., B.Ş.Ö.; Supervision - T.B., B.Ş.Ö.; Funding - T.B., B.Ş.Ö.; Materials - T.B.; Data collection and/or processing - T.B.; Data analysis and/or interpretation - T.B., B.Ş.Ö.; Literature search - B.Ş.Ö.; Writing - B.Ş.Ö.; Critical review ? B.Ş.Ö.