METHODS
We have taken 15 Patients of ca-esophagus for our study from our hospital database. Patients were already
treated 3DCRT plan. Retrospectively additional IMRT and RA plans were created with Planning
target volume (PTV) prescription dose of 50.4Gy in 28 fractions. Prescription dose of 95% of PTV was
chosen for the comparison between three treatment planning strategies. Dose volume histogram was
used to analyze and compare various plan quality index and doses to the organs at risk (OARs).
RESULTS
The D95% of PTV for 3DCRT plan was 47.85±0.78 Gy, which significantly got improved in IMRT
(48.75±0.86 Gy) and RA (49.0±0.64 Gy) plans. The CI value for PTV coverage in 3DCRT was 0.96±0.04;
however, the CI values in IMRT and RA plans were significant at 0.98±0.01 and 0.98±0.03, respectively.
The 3DCRT plan had an HI of 1.09±0.03, which was significantly raised to 1.046±0.26 in the IMRT plan
(p=0.01) and 1.06±0.02 in the RA plan (p=0.03).
CONCLUSION
An IMRT and RA plan gives greater conformity of dose to the PTV in comparison to 3DCRT. Lesser
doses to the surrounding OARs are the major importance of the IMRT and RA planning technique.
Dose escalation is possible with IMRT and RA plans.
Keywords: Ca-esophagus; intensity-modulated radiation therap; plan quality index; RA; three-dimensional conformal radiation therapy
However surrounding organs at risk (OARs) often limits the tumor dose. Planning target volume (PTV) is surrounded by lung, heart, and sometimes abutting to spinal cord. Studies have concluded that the use of modern radiation therapy (RT) techniques is helpful to explore the possibility of dose escalation.
Advances in RT technology have shifted for EC treatment
techniques from three-dimensional conformal
RT (3DCRT) to intensity-modulated RT (IMRT) and
volumetric modulated arc therapy (VMAT) to allow for
the sparing of normal healthy organs such as the lungs,
heart, stomach, liver, kidneys, and spinal cord. IMRT
treatment plan shows superiority to 3DCRT treatment
plan based on dose volume analysis.[
RA produces dose distribution similar to that of
IMRT using a single or multiple arc rotation of the gantry.
Recent research indicated that VMAT is superior
to traditional IMRT in terms of PTV conformity and
dose reduction in OARs.[
The current study's goal is to determine the potential
comparative clinical benefit of the 3DCRT, IMRT,
and RA treatment techniques used in the treatment of
EC, to determine the dosimetrically appropriate technique
in terms of dose to PTV and minimum dose to
OARs, and to determine whether dose escalation is
possible with higher techniques.
After acquiring the CT images of the patients"
gross tumor volume (GTV) and other associated target
volume (TV) were contoured according to the RT
Oncology Group (RTOG) 0436 protocol by the Radiation
Oncologist.[
Dose Prescription and Treatment Objective
Radiotherapy Planning
Different treatment planning strategies were involved
in the treatment of the esophageal cancer patient.
3D-CRT Planning
IMRT Planning
RA Planning
In this study,RA plan with two full arcs and with
single isocenter was generated Arc consists of a clockwise
and a counter-clockwise arc from gantry angles 181.0-179.0 and 179.0-181.0, respectively with collimator
rotation of 15-20°. Collimator rotation was given
to cover the entire tumor and minimize tongue and
groove effect of MLCs during arc rotation.
Plan Evaluation and Statistical Methods
HI
CI
The value of CI to 1 indicates the good dose conformity.
In addition, treatment parameters including
the MU and beam on time (BOT) for each treatment
plan were recorded for evaluation. BOT was defined
as the radiation delivery time and did not incorporate
gantry movement, the patient positioning and imaging
procedures, which was noted while performing quality
assurance.
For the evaluation of target conformation .The
Conformation number (CN)[10] is defined as
CN=[TVRI/TV]*[TVRI/VRI] (3)
TV, TVRI, and VRI represent the treatment volume,
the treatment volume at reference isodose (RI)
of the prescribed dose, and the total volume at RI of
the prescribed dose, respectively. The RI was defined as
95% of PTV prescribed dose. The maximum value for
CN is 1, corresponding to perfect PTV coverage.
To asses overall plan quality of the treatment plans
another parameter named quality factor (QF) was calculated.[
Where the values of the weight factor (Wi) can
be modified between zero and unity for all relatively
weighted indices Xi for a user-defined number of indices
(N) in the UPI set. Here, in our case, we have set all
the weight factor to 1.
A test of significance was required to quantify the
differences between observed parameters of 3DCRT,
IMRT and RA plans. The one-way analysis of variancewas
used to compare dosimetric and volumetric
indices among different plans, and when an overall
significant difference among the groups was observed,
the post hoc Turkey's test was used to determine which
pair-wise comparisons differed. All statistical tests
were done using the IBM Statistical Package for the Social
Sciences (SPSS) software (release 20.0, SPSS Inc.,
Chicago, IL, USA). Statistical significance was defined
as p<0.05.
Each patient was scheduled to receive 1.8 Gy in 28
fractions for a total dose of 50.4 Gy. The planning goal
was to deliver 95% of the prescribed dosage (PD) to
95% of the PTV volume, with no more than 2% of the
PTV volume getting 107% of the PD and a maximum
dose of <110% of the PD. The major aims of the OARs
were defined as: Spinal cord Dmax<45 Gy; and lungV20Gy
<37% and V30Gy <25%. The secondary objectives were
as: Mean dose of lungs <20 Gy; heart V40Gy 50%; and
mean dose of heart <30 Gy.
The LA used to deliver treatment planning was Varian
Vital Beam equipped with the Millennium 120 multileaf
collimator (MLC). Photon beam of energies 6MV,
10MV, and 15MV was used in this study for all cases
treatment planning. There are total 60 pairs of leaves
with 40 leaf pairs in the center and ten pairs on either
side. The projection of center leaf width at isocenter
was 5 mm while the outer leaves were 10 mm. The
maximum travelling speed of MLC was 2.5 cm/s. The
treatment planning system used for the external beam
planning was Eclipse (Version 13.6.23, Varian Medical
System) and final dose calculation was performed
by Anisotropic Analytical Algorithm (AAA, version
13.6.23).Retrospectively new additional IMRT and RA plan were created for each patient. Inverse optimization
was performed by Photon Optimizer (PO) utilized
for both IMRT and RA optimization (version 13.6.23).
3DCRT plans were scheduled to perform into two
phases of treatment, with each phase planned separately
for the same PTV. Phase-I consisted of a giving 36 Gy
dose to PTV with Parallel-opposed, anterior-posterior
(AP), and posterior-anterior (PA) 6 MV, 10MV, and
15MV photon beam, followed by remaining dose of
14.4 Gy in Phase-II, with three photon field. Beam energy
selection was based on patient thickness and PTV
coverage. A beam angle for phase-II planning includes
an AP field and 2 posterior oblique fields at gantry angles
between 100-120° and 240-260° chosen in such a
way to avoid the spinal cord. Appropriate wedge angles
were used in posterior oblique beams to achieve optimum
PTV dose distribution.
This treatment planning strategy involved treating
the PTV with total 54 Gy in 1.8 Gy into 30 fractions.
Inverse optimization was performed with PO which
helps in computing optimal fluence maps from dosevolume
constraints derived from the general planning
objective. Varian leaf motion calculator was used
to calculate actual fluence which is then delivered by
MLCs to deposit the planned dose. Final dose calculation
was performed by AAA algorithm.
The RA is unique treatment technique that allows the
radiation to be delivered to the patient in a single 360°
of gantry rotation that is accurately and efficiently with
varying velocities and positions of the MLCs, dose rate,
and gantry speed. This leads RA being an intensity-modulated
dose distribution. The RA dose optimization is an
aperture-based method, which incorporates MLCs leaf
positions and MU weights as optimization parameters.
The entire gantry rotation is described in the optimization
process by a sequence of 177 control points. The optimization
is based on the PO algorithm and the process
is divided into five discrete multi resolution levels (MRs).
The PO needed less control points to approach the convergent
solution. The dose distribution is calculated during
optimization with a MR dose calculation algorithm.
As per International Commission on Radiological
Units and Measurements (ICRU Report 83, 2010)[
HIwas evaluated as the difference between the dose to
2% (D2) and 98% (D98) of PTV divided by dose to 50%
HI=(D2%-D98%) ? D50% (1)
Where D2%, D98%, and D50% are the minimum dose
delivered to 2%, 98%, and 50% volume of the TV, respectively.
HI of zero indicated homogeneous dose distribution.
A ratio evaluating the coverage of the prescription dose
in treatment plans.

Dmax dose to TV was substantially greater in the
3DCRT (54.03±0.556 Gy) plan than in the IMRT
(52.46±1.031 Gy) and RA (53.2±0.41 Gy) plans, with
p values of (p=0.01) and (p=0.046), respectively. There was no significant difference in Dmax dosage between
IMRT and RA plans (p=0.069) (Table
With 50.59±0.24 Gy, 50.59± 0.04 Gy, and 50.06±0.24
Gy, respectively, the mean dose to PTV in 3DCRT and
RA plans was higher than in IMRT plans. The D95% of
PTV coverage in the 3DCRT plan was 47.85±0.78 Gy,
which was significantly better than the IMRT and RA
plans, which were 48.75±0.86 Gy and 49.0±0.64 Gy,
respectively (Fig.
IMRT: Intensity-modulated radiation therapy; 3DCRT: Three-dimensional conformal radiation therapy.
DVH: Dose volume histogram; PTV: Planning target volume; IMRT: Intensity-modulated radiation therapy; 3DCRT: Three-dimensional
conformal radiation therapy; Gy: Gray.
The CN value for 3DCRT was reported to be 0.56±0.1, which improved significantly to 0.74±0.12 in IMRT (p=0.001) and 0.81±0.05 in RA (p=0.001), respectively. It was found that in 3DCRT, the CI value for PTV coverage was 0.96±0.04, whereas in contrast to IMRT and RA plans, the CI values were significant at 0.98±0.01 and 0.98±0.03. The HI of the TV was 1.09±0.03 in the 3DCRT plan, which was substantially enhanced to 1.046±0.26 in the IMRT plan (p=0.01) and to 1.06±0.02 in the RA plan (p=0.03).
Table
The average heart dosage in 3DCRT was 32.32±9.16 Gy, but the average heart dose in IMRT and the RA plan was 25.08±4.8Gy (p=0.06) and 22.99±6.13 Gy (p=0.015), respectively, with significantly lower radiation dose. The heart's V25Gy was 67.74±23.31 Gy in 3DCRT, but this reduced to 42.83±10.87 Gy (p=0.017) and 34.71±14.97 Gy (p=0.001) in the IMRT and RA plans, respectively. Heart's V30Gy dose in the 3DCRT plan exceeded our clinical dose requirements by a wide margin, but the dose was considerably reduced in the IMRT (p=0.02) and RA (p=0) plans. In 3DCRT, the heart"s V30Gy was 61.5227.06 Gy, whereas it was only 31.697.04 Gy (p=0.002) and 24.2910.43 y (p=0.001) in the IMRT and RA plans, respectively.
In the 3DCRT plan, the average MU required to deliver
a dose of 50.4 Gy was reported to be the lowest
(218.72±7.9). While the average MU required to give
the same dose in IMRT and RA plans was 698.80±179.0
(p=0.001) and 510.98±86.27 (p=0.001), respectively.
BOT was found to be less (0.43±0.14 min) in 3DCRT,
which increases to 1.16±0.29 min and 0.85±0.14 min
in IMRT and RA plans, respectively (Table
In this study, the dosimetric characteristics of
IMRT and RA plans in the treatment of patients with
middle one-third esophageal cancers were studied by
comparing them to 3DCRT plans. IMRT and RA plans
showed the increased dose conformity homogeneity
and PTV coverage when compared to 3DCRT (Fig.
In RA and IMRT plans, the QF is higher than in
3DCRT. This could be due to better CI, HI, and CN
values reported in IMRT and RA plans. IMRT and RA
plans have the ability to cover the PTV with prescribed
dose and simultaneously they deliver high dose gradient
outside the PTV. This results in high conformity
to the PTV.IMRT and RA plans reduced the amount
of normal tissues irradiated with a high dosage while
increasing the volume of normal tissues treated with
a low dose. The study of Vivekanandan et al.[
Pulmonary toxicity is a major potential adverse
effect of treatment for esophageal cancer. As a result,
the dose constraint to bilateral lung is the most important
factor to consider when evaluating treatment
regimens. Numerous studies had been published having
contradicting parameters for predicting pulmonary
toxicity,[
Numerous studies have found significant radiationinduced
heart damage when the heart received more
over 40Gy, and that lowering V40Gy was important in reducing
heart toxicities. Pao et al.[
The spinal cord is a dose-restricted organ in the
case of EC, and it should be kept out of the field of
radiation when the neck, thorax, abdomen, and pelvis
are all being treated. Kirkpatrick et al.[
Scattered dose is directly related to MU number.
In comparison to IMRT and RA, 3DCRT plans
had the lowest MU in our study. When compared to
3DCRT, IMRT, and RA treatment plans utilize up to
3 times as many MUs (Fig.
3DCRT: Three-dimensional conformal radiation therapy;
IMRT: Intensity-modulated radiation therapy.
3DCRT: Three-dimensional conformal radiation therapy;
IMRT: Intensity-modulated radiation therapy; Gy: Gray.
3DCRT: Three-dimensional conformal radiation therapy;
IMRT: Intensity-modulated radiation therapy; DVH: Dose
volume histogram.
MU: Number of monitor unit; 3DCRT: Three-dimensional
conformal radiation therapy; IMRT: Intensity-modulated
radiation therapy.
Peer-review: Externally peer-reviewed.
Conflict of Interest: All authors declared no conflict of interest.
Financial Support: None declared.
Authorship contributions: Concept - D.K.S., G.G., S.S., S.Y., N.P.; Design - D.K.S., S.S., S.Y.; Supervision - S.S., D.K.S., S.Y., N.P.; Funding - None; Materials - None; Data collection and/or processing - D.K.S., S.S., S.Y.; Data analysis and/or interpretation - D.K.S., S.S., S.Y., N.P.; Literature search - D.K.S, G.G., S.S., S.Y.; Writing - D.K.S., S.Y.; Critical review - D.K.S., G.G., S.S., S.Y., N.P.