Keywords: Patient positioning; radiotherapy; surface guidance
QA PROGRAM FOR SGRT
Commissioning
Periodic QA Program
QA Phantoms for SGRT
Challenges in SGRT QA
Acceptance Test
The acceptance process need to include all required tests
including static/dynamic localization accuracy, spatial
reproducibility and drift to illustrate the safe operation
and proper functionality of the SGRT system with the
integrated treatment or simulation platform. In most
cases, the acceptance test document is provided by the
vendor and it may not include all necessary tests that
need to be checked. However, it is important to keep in
mind that the acceptance procedure is an integral part
of the purchasing process to ensure whether the product
or solution meet the clinical need, or not. Therefore,
primary responsible person, generally qualified medical
physicist expert, needs to be familiar with the fundamental
or basic tests recommended in the commissioning
and if these tests are not included in the vendor's
acceptance documents, it is generally recommended to
negotiate with the vendor to perform these tests during
acceptance. According to AAPM TG-302, vendor's
recommendation and other AAPM reports such as TG-
142, TG-147, and TG-76 need to be followed together
for checking the localization accuracy and reproducibility
of the system. In addition, safe operation and proper
functionality of the system with all other unit interface,
including imaging system (if necessary), treatment machine,
treatment planning system, data transfer and information
system, and need to be validated as described
in Table
The commissioning of the SGRT system is a substantial
part of the comprehensive QA program before implementing
it into clinical practice. This part also includes
measuring the system accuracy/precision and determining
system limitations for all clinically relevant
scenarios. Since the commissioning data are accepted
as a reference for future measurement, all tests need to
be reproducible to assess the consistency of the system
performance over the period of time for periodical QA
tests or for later measurements after maintenance and
repair of the system. In addition, according to AAPM
TG-147 recommendation, commissioning test need to be repeated in case of special situation, ranging from
major upgrade and power outages to earthquake and
building vibration, to check the stability of the system
before using it in clinical practice. All suggested parameters
for system commissioning in AAPM TG-302
and ESTRO-ACROP guidelines are summarized in
Table
The main goal of the periodical QA program is to ensure
about the stability of the system over a period of
time (e.g., daily, weekly, monthly, and annually) and
to catch the unexpected errors or changes in system
performance due to the many factors such as component
failure, machine malfunction or aging of the
system component. ESRTO-ACROP guideline also
recommended to start with a higher frequency and
higher number of tests until the RT team feel more
comfortable about the stability of the system based on
the test outcome preferably including a failure modes
and effective analysis specific to the clinic. In addition,
ESTRO-ACROP guideline reported the list of failure
modes and potential errors in SGRT workflows with
possible solutions. Similar to acceptance and commissioning
part, ESTRO-ACROP guideline provides more
comprehensive periodic QA program compared to
AAPM TG-302 and TG-147 recommendations as presented
in Table
SGRT requires dedicated QA phantoms with specific
properties (e.g., color, reflectivity, texture, and topography)
that make it accurately trackable. Although some
commercially available SGRT systems allow the user
to change imaging parameters (e.g., camera light and
exposure time) for capturing surface information from
the bodies/phantoms with variety skin/surface tones,
opaque/matte and light colored phantoms yields the
best monitoring results during QA due to the better reflection
characteristic for the projected light pattern. In
fact, the use of SGRT system in variety skin tones, especially
in case of dark skin, is still one of the challenging
issues to consider in clinical practice. However, ESTROACROP
guidelines recommended to check localization
accuracy of the SGRT system with both light- and darktoned
phantoms when it is possible, especially in clinics
where a larger proportion of patients with darker skin tones are treated. In addition, it needs to be taken in to
account that if the surface of the phantom is shiny, it
may also cause numerous or unwanted reflection pattern
of the projected light. Therefore, in case of necessity,
it is generally recommended to cover the phantom
surface with a paint coat or light colored tape. In addition
to color and reflectivity properties, topography
and texture of the QA phantom may significantly affect
the result of the QA tests. Indeed, in case of insufficient
topography, it is difficult to discern position or motion
of the phantom during the check of localization accuracy
of SGRT system. To overcome this issue, vendors
provide dedicated phantoms that mimic anatomical
surfaces such as the head, leg, or breast. In many clinics,
homemade Styrofoam phantom with a different topography
is also used as an inexpensive way of 3D surface
phantom for SGRT. However, we need to be careful
that Styrofoam with expanded polystyrene beads may
cause uncertainties due to the abundance of texture and
the projected light pattern may not be identified correctly.
Therefore, smoot foam phantoms satisfying the
outlined recommendation in both ESTRO-ACROP and
AAPM guidelines can be also good alternative to commercially
available phantoms. Several types of commercially
available phantoms were also demonstrated in
AAPM TG-302 and ESTRO-ACROP guidelines.
As also defined in AAPM TG-302, there are still several
major issues that cause in uncertainties during both
QA and clinical practice of SGRT. For instance, the use
of DICOM based surface structure generated from CT
imaging is considered as the one of the challenging issue
for accurate localization of the phantom/body. In fact,
many parameters (e.g., CT voxel size, scan speed, respiratory
phase effect for moving phantom/surface, Hounsfield
unit threshold for surface segmentation, and image
quality) can significantly affect the topography of
reference body surface ant it may cause a systematic bias
during localization. Similarly, the size and the shape of
the selected region-of-interest for surface tracking can
also affect the response of the system during QA. In
addition to these parameters, the tracking accuracy of
the SGRT system can decreases when the component of
treatment unit (e.g. gantry head and kV imaging arms)
occlude the SGRT cameras, especially in non-coplanar
treatment techniques with couch angle. Therefore, all
these parameters need to be checked for different scenarios
to evaluate the impact of defined issues on the
tracking and localization accuracy of the implemented
SGRT system before using in clinical practice.
Peer-review: Externally peer-reviewed.
Conflict of Interest: I have no conflict of interest.
Financial Support: None declared.