METHODS
The study included 19 patients with gynecologic cancer who received adjuvant radiotherapy. Nineteen
planning computed tomography scans and 171 megavoltage computed tomography (MVCT) scans were
evaluated retrospectively. All structures were re-contoured on MVCT images using MIM software. The
isodoses obtained from the reference plan were loaded onto the contoured MVCT images, and dosevolume
histogram-based analysis was performed.
RESULTS
Analyses showed that patients with a bladder volume of ?300 cc at the time of planning had lower bladder
volumes (p=0.03) and higher mean bladder doses (p=0.04) during treatment compared with planning.
Bladder volume deviation was found to be significantly higher in patients with a bladder volume of ?300 cc
at the time of planning (p=0.002). The mean treatment CTV V95% values of patients with planning bladder
volumes of <150 cc, 150?300 cc, and ?300 cc were 97.2%, 98.7%, and 98.9%, respectively (p=0.005).
CONCLUSION
Bladder and rectum filling have a critical impact on patients receiving adjuvant radiotherapy for gynecologic
cancer. For optimal results, the bladder volume should be planned between 150 and 300 cc.
Keywords: Adjuvant; bladder; filling; gynecologic cancer; radiotherapy; rectum
Patient Preparation and Computed Tomography
(CT) Simulation Scanning
CT images were obtained on a GE Discovery RT simulation
device (GE Healthcare, USA). According to our
clinic protocol, these patients were advised to empty
their rectum and fill their bladder before CT imaging.
Patient stabilization was achieved using a T-board and
a knee and feet positioner. Patients were instructed to
drink 1000 mL of water within 30 minutes after emptying
their bladders, followed by CT simulation. If patients"
bladders were not adequately filled, they were
advised to drink an additional 500 mL of water and
undergo simulation after 15 minutes. In cases where patients" bladders were excessively full, emptying the
bladder and restarting the protocol were recommended.
There was no specific protocol for emptying the rectum; if patients had regular bowel movements in the morning, no additional treatment was recommended. However, if they were constipated or a properly empty rectum could not be obtained, laxative use was prescribed. Following this preparation, a CT scan with a slice thickness of 2.5 mm was performed with the patient positioned supine and the head in.
Contouring
The nodal CTV was created to include the obturator,
internal-external iliac lymph nodes, and presacral
lymph nodes, covered according to stage and histopathology.
In the presence of metastases at the level of the
common iliac or para-aortic lymph nodes, the paraaortic
lymph nodes were included in the nodal CTV.
The planning target volume (PTV) was created by giving
a 7-mm margin to the nodal and vaginal cuff CTVs.
Planning
Megavoltage CT (MVCT) Scanning
Recontouring on MVCT Images
Dosimetric Evaluation
The cut-off point for the rectum AP diameter was
chosen as 4.2 cm, which was the mean value for all patients,
while the cut-off point for the bladder volume was
set at 300 cc, based on previously published literature.[
Statistical Analysis
The difference in parameters between the groups
formed according to bladder and rectum size was analyzed
using an independent t-test and one-way analysis of
variance (ANOVA). Furthermore, a post-hoc analysis was
conducted using Tukey"s test to determine whether significant
differences existed in the treatment CTV V95%
values among subgroups classified by planning bladder
volume. The deviation of bladder volume from planning
bladder volume was demonstrated using scatter plots.
Pearson"s correlation coefficient was used to assess correlations
between the dimension and dose changes. P-values
of <0.05 were considered statistically significant. All statistical
analyses were performed using IBM SPSS Statistics
version 24.0 software (Armonk, NY: IBM Corp. 2016).
The target volumes and OAR were delineated according
to the Radiation Therapy Oncology Group (RTOG)
consensus panel atlas.[
Treatment plans were created using the TomoTherapy
HDA IDMS Precision Planning System software (version
2.0.1.1; Accuray, Madison, Wisconsin, USA). In
the plans, the prescribed dose was intended to cover
95% of the PTV. For the jaw width, 2.5 and 5 cm were
preferred in the dynamic mode, and the pitch factor
was chosen as 0.436 and 0.433, respectively. For the
modulation factor, values between 2-3 were used. In
treatment plans, it was aimed that the 40 Gy dose (V40)
received by the bladder was less than 40%, and the 50
Gy dose (V50) received by the rectum was less than
50%. Additionally, it was aimed that the volume of the
intestines receiving 45 Gy should not exceed 195 cc.
Preliminary preparation was made for patients in accordance
with the treatment planning CT before the treatment.
Before each treatment, 3D MVCT imaging was
performed to check the patient"s position, bladder, and
rectum. Following the positioning of the patient, the radiation therapist evaluated the bladder and rectal filling.
If the volumetric alterations in these organs did not
result in the displacement of CTV beyond the PTV, the
treatment proceeded. If the fullness of the bladder and
the empty rectum was not acceptable, the patient"s pretreatment
preparation was redone as described in the CT
simulation patient preparation section. The MVCT scan
length was determined to include the PTV to shorten
the scan time and reduce the dose the patient would receive
from the scan. MVCT scanning parameters were
chosen as coarse for pitch and 3 mm for the interval.
A total of 190 CT images (including 19 planning CT and
171 MVCT images) were evaluated in the study. Nine
MVCT images for each patient, obtained at 3-day intervals
starting from the first day of the patient"s treatment,
were used. These images were transferred to MIM software
running with Accuray TomoTherapy Precision
TPS, and automatic deformable fusions were made on
the planning CT image. The system automatically created
CTV, bladder, and rectum in all images. Subsequently,
radiation oncologists corrected these constructs in
all MVCT images. The simulation CT plan is referred to
as planning CT, and plans made using MVCT images in
the MIM are named treatment CT.
Each patient"s reference plan was recalculated using
MIM software on all MVCT images where new contours
were created. Dimensions and doses during treatment
were considered the mean value of dimensions,
and doses were obtained from 9 MVCT scans for each
patient. In the study, besides CTV, dose changes in the
bladder and rectum as OAR were evaluated. The dose
covering 95% (V95%) was evaluated for CTV. Bladder
mean dose and the percentage of bladder volume (%)
that received a 40 Gy dose (V40) were examined. For
the rectum, the volume (cc) covered by the 40 Gy dose
(V40) and AP diameter were evaluated.
Descriptive statistics were performed to present baseline
characteristics. The Kolmogorov-Smirnov and
Shapiro-Wilk tests were employed to assess the normality
of the data distribution. Parametric tests were
applied as the data satisfied the assumption of normality
based on the results of these tests. The planning and
treatment parameters were compared using the paired
t-test. Repeated measures ANOVA was used to analyze
the difference between treatment fractions.
Comparisons of bladder volume, mean dose,
and V40 values between planning and treatment are
given in Table
Assessment of the plans based on MVCT images
revealed no statistically significant differences in bladder
volume, mean dose, and V40 between treatment
fractions (p=0.16, p=0.51, and p=0.79, respectively).
The deviation of the bladder volume from planning to
treatment for each evaluated fraction was obtained, and
the mean deviation volume was calculated for each patient.
Patients with planning bladder volumes of <300
cc and ≥300 cc were compared for bladder volume
deviations. Notably, patients with a bladder volume of
?300 cc during planning exhibited significantly higher
bladder volume deviations (95% CI: [45.10-161.40 cc];
p=0.002) (Table
Table
When all patients were included in the analysis, no statistically significant difference was found between planning and treatment regarding rectum AP diameter and V40. Patients with a rectum AP diameter of <4.20 cm at planning had higher rectum AP diameters during treatment compared with planning (95% CI: [0.05?0.78 cm]; p=0.02). Patients with a rectum AP diameter of ≥4.20 cm at planning had lower rectum V40 values during treatment compared with planning (95% CI: [1.51-8.52 cc]; p=0.01).
Assessment of the plans based on MVCT images revealed no statistically significant difference in rectum AP diameter and V40 between treatment fractions (p=0.75 and p=0.24, respectively). The mean CTV V95% value for all patients at planning and treatment was 100% and 98.4%, respectively (95% CI: [1.02- 2.07]; p=0.001). The mean treatment CTV V95% values of patients with planning bladder volumes of <150 cc, 150-300 cc, and ≥300 cc were 97.2%, 98.7%, and 98.9%, respectively (p=0.005).
Post hoc analysis revealed lower treatment CTV
V95% for a planning bladder volume of <150 cc
compared with 150-300 cc and ≥300 cc (p=0.02 and
p=0.004, respectively). However, no significant difference
emerged between 150-300 cc and ≥300 cc regarding
CTV V95% (p=0.89). In addition, planning
bladder longitudinal diameter and planning bladder
volume correlated with treatment CTV V95% (r=0.53,
p=0.01 and r=0.48, p=0.03, respectively). The change in bladder volumes and CTV contours between two
treatment fractions of a patient is shown in Figure
CTV: Clinical target volume.
On the other hand, when the treatment CTV V95% values of patients with planning rectum AP diameters of <4.2 cm and ≥4.2 cm were compared, no statistically significant difference was identified (p=0.66).
Correlation analysis was also performed by taking the differences between the treatment and planning data. Changes in bladder volume were negatively correlated with changes in bladder mean dose (r=-0.77, p<0.001), bladder V40 (r=-0.72, p=0.001), and CTV V95% (r=-0.53, p=0.01). Changes in rectum AP diameter were positively correlated with changes in rectum V40 (r=0.57, p=0.01). There was no significant correlation between changes in rectum AP diameter and changes in CTV V95%.
Although the bladder filling protocol was implemented
in our study, significant differences in bladder
volume were observed among the patients. Chan
et al.[
We divided the sample into two groups based on
planning bladder volume (cut-off 300 cc) and found
that the group with greater bladder volumes at planning
CT had lower reproducibility during treatment.
In agreement with our findings, Eminowicz et al.[
Our results showed that bladder volumes above
300 cc in the initial CT simulation led to higher mean
doses and V40 of the bladder during treatment compared
with planning. Greater bladder volumes in the
simulation cause an increase in the doses received by
the bladder as an OAR and can lead to alterations in target coverage by changing the position of the uterus
or vaginal cuff. Okamoto et al.[
Therefore, it is recommended to maintain planning
bladder volumes within the range of 150-300 cc to
minimize deviations during treatment and reduce radiation
doses to nearby organs.
Shah et al.[
In addition, we showed that the planning bladder
longitudinal diameter and bladder volume were correlated
with treatment CTV V95%. Changes in bladder
longitudinal diameter and volume affect the location
of the bowel. In patients with short longitudinal diameters
of the bladder, the bowel is displaced towards the
lower parts of the pelvis, which results in shorter contouring
of CTV. This smaller target delineation causes
decreased CTV doses in treatment.
To monitor bladder volume, patients should be
subjected to a standardized protocol; however, it has
been observed that the same protocol can lead to different
bladder volume fillings in different patients.
Therefore, during the initial days of treatment, patients"
adherence to protocols should be carefully
monitored, and personalized protocols, such as shortening
the waiting period or increasing fluid intake,
should be established accordingly.
Regarding rectal filling, a systematic review including
five studies comprising 103 patients undergoing definitive
radiotherapy demonstrated that rectal filling had a
more pronounced effect on the motion of the cervix and
the upper part of the vagina compared with the uterus.
Also, a significant correlation between the rectal volume
and the AP shift of the CTV has been shown.[
Agrawal et al.[
Even though we found no correlation, many studies
have demonstrated a relation between rectal filling
and CTV motion. Especially in clinics where IGRT
is not used, patients should be instructed to empty
their rectum before both simulation and treatment to
reduce variations.
Daily IGRT ensured consistency between the actual
CTV doses and the prescribed CTV dose in our study.
As mentioned in the method, if the changes in the
bladder and rectum before treatment are substantial
enough to cause the target to exceed the pre-defined
PTV, the patient"s preparation is re-done. This clinical
protocol leads to better compatibility of CTV coverage
and OAR doses with planning.
Limitations
It should be noted that the treatment period in the
study ranged from 25 to 28 days, but nine CT scans
were selected for each patient to analyze. To enhance
the strength of the study, it would be valuable to include
evaluations based on data collected throughout
the entire treatment course.
IGRT is a strong tool to demonstrate the displacements of target volumes and OAR. Daily use of IGRT helps increase the compatibility of treatment doses with planned doses, and it will also help calculate the cumulative dose accurately for patients receiving adjuvant radiotherapy and brachytherapy. Dose calculations based on IGRT images can be helpful in deciding offline adaptive planning. Therefore, if possible, the implementation of daily IGRT should be put into practice.
Ethics Committee Approval: The study was approved by the Akdeniz University Faculty of Medicine Clinical Research Ethics Committee (no: 774, date: 28/12/2022).
Authorship contributions: Concept - T.K., Y.B.; Design - R.A.A., Y.B.; Supervision - E.A., M.K.; Funding - N.T., A.F.K.; Materials - Y.B., İ.H.S.; Data collection and/or processing - R.A.A., M.K., E.A.; Data analysis and/or interpretation - R.A.A., Y.B., E.A.; Literature search - T.K., A.F.K.; Writing - T.K., R.A.A., E.A., M.K., Y.B., N.T., İ.H.S., A.F.K.; Critical review - T.K., R.A.A., E.A., M.K., Y.B., N.T., İ.H.S., A.F.K.
Conflict of Interest: All authors declared no conflict of interest.
Use of AI for Writing Assistance: No AI technologies utilized.
Financial Support: None declared.
Peer-review: Externally peer-reviewed.