Guide For Quality Control and Quality Assurance Tests in Tomotherapy/Radixact Systems

Evren Ozan GÖKSEL1, Songül BARLAZ US3, Yücel AKDENIZ4, Recep BOZCA5, Yelda ELÇIM6, Zerrin GANI7, Gülay GÜRAY8, Nazan DEDE KARAKOÇ9, Nilgül NALBANT10, Turgay TOKSOY11, Abdullah YEŞIL12
1 Acıbadem Mehmet Ali Aydınlar Üniversitesi, SHMYO, Radyoterapi Programı, İstanbul-Türkiye
2 Acıbadem Altunizade Hastanesi, Radyasyon Onkolojisi Kliniği, İstanbul-Türkiye
3 Mersin Üniversitesi Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, Mersin-Türkiye
4 Acıbadem Adana Hastanesi, Radyasyon Onkolojisi Kliniği, Adana-Türkiye
5 Başkent Üniversitesi, Adana Dr. Turgut Noyan Araştırma ve Uygulama Merkezi, Radyasyon Onkolojisi Anabilim Dalı, Adana-Türkiye
6 Gülhane Eğitim ve Araştırma Hastanesi, Radyasyon Onkolojisi Kliniği, İstanbul-Türkiye
7 Ankara Bilkent Şehir Hastanesi, Radyasyon Onkoloji Kliniği, Ankara-Türkiye
8 Medikal Park Bahçelievler Hastanesi, Radyasyon Onkolojisi Kliniği, İstanbul-Türkiye
9 Kartal Dr. Lütfi Kırdar Şehir Hastanesi, Radyasyon Onkolojisi Kliniği, İstanbul-Türkiye
10 Acıbadem Atakent Hastanesi, Radyasyon Onkolojisi Kliniği, İstanbul-Türkiye
11 Ali Osman Sönmez Onkoloji Hastanesi, Radyasyon Onkolojisi Kliniği, Bursa-Türkiye
12 Medicana Bursa Hastanesi, Radyasyon Onkolojisi Kliniği, Bursa-Türkiye
Published: 2024 DOI: 10.5505/tjo.2024.003 Article ID: 1491
Abstract
The MFD-GG003 task group has prepared a quality control and assurance report for Tomotherapy/
Radixact devices. This published report summarizes the main test contents and application methods of
the primary report. The purpose of this report is to standardize the quality control tests and methods
recommended by manufacturers and international guidelines for Tomotherapy/Radixact devices and
to provide recommendations for medical physicists in radiation oncology clinics to establish safe and
effective quality assurance programs. The quality control tests are categorized under headings such as
mechanical tests, water phantom tests, film tests, solid phantom tests, imaging component tests, and
treatment planning system tests. For each test, the purpose, procedure, recommended frequency, and
tolerance range are specified. Additionally, the recommended tests, along with their tolerance values,
are listed in tables for daily, monthly, and annual intervals. The suggested test frequencies and tolerance
values were determined based on international references, manufacturer recommendations, and the
clinical experience of the MFD-GG003 task group members. However, when establishing a quality assurance
program for a clinic, medical physicists must make appropriate decisions for their own clinics
by considering device characteristics and the treatment techniques applied, especially when modifying
or determining tolerance values and test frequencies.

Keywords: Quality assurance; radixact; tomotherapy