METHODS
Co-60 teletherapy equipment, Orion 4 (MV), Saturn 42 (6-15 MV) high energy x-rays were used. Measurements were made for different field sizes and source skin distances by using Marcus–type parallel plate ion chamber at water equivalent solid phantom, at surface and build-up region. Measured surface and build-up doses were installed to CMS XiO treatment planning system.
RESULTS
As a result of this study, the surface dose increased linearly as the field size increased and the maximum dose point was moved toward the surface by 1-2 mm as predicted. Surface doses for a 10x10 cm field size are 24.3%, 17.4%, 12.6% and 10.2% respectively for Co-60, 4MV, 6MV and 15MV. With the increase in skin source distance, a decrease in surface doses is observed.
CONCLUSION
The use of correct information is achieved by downloading surface doses measured for different field sizes to the treatment planning system.
Keywords: Build-up region; high energy photons; surface dose; radiotherapy dosage