Comparison of theoretical and experimental values of TRS 398 absorbed dose to water calibration factors and beam quality dependence factors

Hilal ACAR
1 İstanbul Üniversitesi Onkoloji Enstitüsü, Klinik Onkoloji ve Radyasyon Onkolojisi Ana Bilim Dalları, İstanbul
Published: 2008 Pubmed: 77010 Article ID: 692
Abstract
OBJECTIVES

In this study, experimentally and theoretically determined
absorbed dose to water calibration factors were compared.
We used 6 MV photon and 12 MeV electron energies to compare
experimentally and theoretically determined beam quality
dependence factors.


METHODS

Measurements were made at SSD=100 cm distance (L IN
A C), 10x10 cm standard field size and reference depth.
Five different cylindrical and three parallel plate chambers
were used.


RESULTS

For absorbed dose to water calibration factor, the max difference
between theoretical and experimental values was 2.8%
for Marcus chamber. Beam quality dependence factors for 6
MV theoretical values of PTW 30001 (a) (-0.816%), NE 2571
(-0.1107%) were greater than their experimental values. For
12 MeV, the difference was 1.2371% for PTW 30001 (a),
1.6774% for PTW 30001 (b), 0.8758% for NE 2571, 0.8363%
for NE 2581, 0.6071% for PTW 30002, and 0.6245% for
Marcus ion chamber.


CONCLUSION

As the protocol suggests, obtaining absorbed dose to water
calibration factor directly from SSDL produces more accurate
results.

Keywords: Absorbed dose to water calibration factor; beam quality dependence factors; TRS 398