Turkish Journal of Oncology Turkish Society for Radiation Oncology
Article Open Access Volume 24 · Issue 1 · 2009 pp. 015–020

The effect of stopping power ratios of different protocols on %DD curve

Hilal ACAR
1 İstanbul Üniversitesi Onkoloji Enstitüsü, Klinik Onkoloji Anabilim Dalı, Radyasyon Onkolojisi Bilim Dalı
Published: 2009 Pubmed: 82021 Article ID: 713
Abstract
OBJECTIVES

The effects of the stopping power ratios of different protocols
on %DD curve were investigated in this study.


METHODS

Percentage ionization curve was measured at SSD=95 cm
using 10x10 standard cone and 6, 8, 10, 12, 15 and 18 MeV
electron energies. Stopping power ratios at each depth were
determined as recommended by the three protocols. %DD
curves were then found and compared. The stopping power
ratios of the different protocols were determined and plotted
against R50.


RESULTS

The stopping power ratios of the different protocols did have
an effect on %DD curves. The clearest difference was seen at
the surface. The maximum difference was 1.45% at 10 MeV,
while the minimum difference was 0.09% at 6 MeV.


CONCLUSION
Stopping power ratios as found using realistic electron beams
provide more accurate surface doses, as reported in the literature.

Keywords: Stopping power ratios; electron beams; clinical protocols/ standards; IAEA protocols