The effect of segment numbers on dosimetric verification of IMRT fields obtained by the static IMRT (step and shoot) method

Hediye ACUN1
1 Harran Üniversitesi Tıp Fakültesi, Biyofizik Anabilim Dalı, Şanlıurfa
2 İstanbul Üniversitesi Onkoloji Enstitüsü, Radyasyon Onkolojisi Kliniği, İstanbul
3 İstanbul Üniversitesi İstanbul Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, İstanbul
Published: 2011 Pubmed: 98712 Article ID: 769
Abstract
OBJECTIVES

The goal of this work was to investigate the effect of segment
number on the verification of intensity-modulated radiotherapy
(IMRT) dose distributions measured with 2D array and
calculated by treatment-planning system (TPS).


METHODS

Five cm depth-fluence maps calculated by TPS and measured
by 2D array were compared using PTW Verisoft program
for the IMRT plans of 16 prostate patients. The percentage
of points passing through the (g) evaluation was found for
different distance to agreement (DTA) and dose difference
(DD)%.


RESULTS

When coronal fluence maps were compared, the percentage
of points passing through the (g) evaluation was obtained as
98.45 (95.5-100) for the low segment group and as 86.75 (73-
94) for the high segment group for 3 mmDTA/3% DD.


CONCLUSION

The percentage of points passing through (g) evaluation was
found lower for the high segment group than for the low segment
group. To minimize this uncertainty in the IMRT verification,
the use of a large number of segments should be avoided
to the extent possible.

Keywords: Intensity-modulated radiotherapy (IMRT); dosimetric verification; 2D array; gamma index