Turkish Journal of Oncology Turkish Society for Radiation Oncology
Article Open Access Volume 27 · Issue 1 · 2012 pp. 024–028

Calibration of gamma rays with LiF thermoluminescence dosimeters

Mustafa Tarkan AKSÖZEN, Kadir YARAY, Selahattin MENTEŞ, Mete GÜNDOĞ, Celalettin EROĞLU
1 Erciyes Üniversitesi Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, Kayseri
Published: 2012 Article ID: 802
Abstract
OBJECTIVES

We aimed in this study to calibrate lithium fluoride thermoluminescence
dosimeters (TLDs), alloyed with magnesium and
titanium, by using gamma rays.


METHODS

168 application-naïve TLDs were annealed and then calibrated
in a TLD-reader system at a sensitivity level of ±1% after
achieving a saturation level by gamma irradiation. Five
selected TLDs were placed in the center of the reference area,
formed by using Rando phantom, and irradiated at the dose of
180 cGy with 6 MV to measure the accuracy of a given dose.


RESULTS

The doses exposed by the TLD were determined with a reader
system as follows: 182.1, 181.0, 180.3, 180.3, and 180.0 cGy.


CONCLUSION

The exposed doses were in the range of ±1% in all TLDs except
TLD#1. The error in the dose measured for TLD#1 was
found as +0.6%. This could be attributed to incomplete release
of energy traps due to the complicated annealing process, temperature
and humidity of ambient and lack of sterilization.

Keywords: Lithium fluoride; rando phantom; thermoluminescence dosimeter