Dosimetric verification of prowess panter treatment planning system and the evaluation of the clinical acceptance

Ayşegül ÜNAL KARABEY1
1 Aktif Care Tıbbi Cihazlar Ltd. Şti., İstanbul
2 İstanbul Üniversitesi Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, İstanbul
Published: 2012 Pubmed: 112700 Article ID: 831
Abstract
OBJECTIVES

In this study, the aim was to perform quality control of the Prowess
Panther treatment planning system by comparing with
the measurement data.


METHODS

For dosimetric measurements, Iba brand cc13 and FC-65P
0.6cc model (Scanditronix, Wellhöfer, Germany), the cylindrical
ion chambers, and Iba brand dose1 electrometer model,
RW3 solid water phantom and RFA-300 water phantom, were
used. Measurements were performed with Elekta Synergy linear
accelerator. RFA-300 Wellhöfer profile measurements were
taken using a water phantom in the wedge and open field depth
dose, off-axis ratio, and different x, y and diagonal directions.
Output measurements were performed with RW3 solid water.


RESULTS

Open and wedge field percentage depth dose (PDD) and profiles
of the treatment planning system were compared with
measurements at Dmax, 5 cm, 10 cm, and 15 cm depths. The
results were within the boundaries of acceptable BJR and
found less than 3%.


CONCLUSION

Based on the quality control results, the Prowess Panther v5.01
Planning System was within the boundaries of acceptable accuracy
in dosimetry.

Keywords: Leaf transmission; profile; prowess panther treatment planning system; output; percent depth dose