Midplane dose determination and verification of calculated doses in total body irradiation

Özlem ÖZDEMİR1
1 Pamukkale Üniversitesi Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, Denizli
2 Ege Üniversitesi Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, İzmir
Published: 2013 Article ID: 858
Abstract
OBJECTIVES

To compare calculated and measured doses for different regions
of anthropomorphic phantom and patients using ion
chamber and thermoluminescence dosimetry (TLD) for total
body irradiation.


METHODS

Measurements were done for lateral fields with 6 MV, gantry
82º, 40x40 cm2 field and 400 cm source-axis distance (SAD).
Entrance-exit and midline doses were measured on anthropomorphic
phantom by TLD and entrance-exit doses were measured
by TLD and ion chamber on patients.


RESULTS

For anthropomorphic phantom measurements differences between
calculated and measured entrance-exit doses of head,
neck, shoulder, lung and thick pelvis were 0.8%, 2.7%, 26.4%,
4.4% and 4.9% and for midline doses were 1.6%, 1.6%, 6.3%,
-1.4% and 7.4% respectively. For patients; TLD differences
were within -4.13% ile 6.7%, -3.3% ile 3.9%, 5.1% ile 16.6%,
-7.8% ile 2.4%, and 3.6% ile 7.1% respectively. For thick pelvis
measurements with ion chamber differences were within
%0.1-1.9.


CONCLUSION

Total body irradiation is being applied in limit values in our
clinic.

Keywords: Total body irradiation, in-vivo dosimetry, thermoluminescence dosimetry