Turkish Journal of Oncology Turkish Society for Radiation Oncology
Article Open Access Volume 39 · 2024 pp. 113–122

Guide for Quality Control and Quality Assurance Tests in Image-guided Radiotherapy Systems

Yağız YEDEKÇİ1, Gizem BAKICIERLER2, Bora SINDIR2, Esin GÜNDEM3, Esil KARA4, Sümeyra CAN5, Taha ERDOĞAN6, Osman Vefa GÜL7, Çağatay Recep ÖZBAY8
1 Hacettepe Üniversitesi, Radyasyon Onkolojisi Anabilim Dalı, Ankara-Türkiye
2 Celal Bayar Üniversitesi, Radyasyon Onkolojisi Anabilim Dalı, Manisa-Türkiye
3 Sağlık Bilimleri Üniversitesi, Gülhane Eğitim ve Araştırma Hastanesi, Radyasyon Onkolojisi Anabilim Dalı, Ankara-Türkiye
4 Etlik Şehir Hastanesi, Radyasyon Onkolojisi Anabilim Dalı, Ankara-Türkiye
5 Başakşehir Çam ve Sakura Şehir Hastanesi, Radyasyon Onkolojisi Anabilim Dalı, İstanbul-Türkiye
6 Afyonkarahisar Sağlık Bilimleri Üniversitesi, Radyasyon Onkolojisi Anabilim Dalı, Afyonkarahisar-Türkiye
7 Selçuk Üniversitesi, Radyasyon Onkolojisi Anabilim Dalı, Konya-Türkiye
8 Acıbadem İzmir Hastanesi, Radyasyon Onkolojisi Kliniği, İzmir-Türkiye
Published: 2024 DOI: 10.5505/tjo.2024.008 Article ID: 1496
Abstract
This report systematically addresses quality control and assurance tests aimed at enhancing the safety and
efficacy of Image-Guided Radiation Therapy (IGRT). IGRT utilizes advanced imaging systems to accurately
delineate tumor volumes, minimizing uncertainties caused by anatomical changes during treatment. The study
encompasses quality control tests for 2D, 3D, and 4D imaging systems integrated with linear accelerators, as
well as surface and respiratory tracking systems. The tests are categorized into three main groups: radiographic imaging systems, surface tracking systems, and respiratory tracking systems. The report details procedures and
tolerance values for critical parameters such as geometric accuracy, image quality, and safety tests. Additionally,
quality control methods for respiratory-gated radiation therapy systems and 4D-CBCT modules are presented.
The report emphasizes the importance of standardizing quality control protocols to ensure the accurate
and reliable implementation of IGRT and aims to serve as a guide for newly established clinics and medical
physicists, providing a reference for developing and implementing their own quality control procedures.

Keywords: 4D-CBCT; geometric accuracy; IGRT; quality control; respiratory tracking